During a multipoint dissolution test, there are two common approaches after withdrawing each sample:
- Replace the aliquot with fresh dissolution medium. Or
- Do not replenish the medium and apply the appropriate volume correction.
At first glance, replenishment may appear to be the cleaner option.
Withdraw 5 mL --> Add 5 mL = Maintain the original vessel volume.
However, dissolution occurs within a highly sensitive hydrodynamic environment
Adding fresh medium during an ongoing test can disturb the vessel hydrodynamics, disintegrating/ eroding mass, settled particles etc. This risk becomes more relevant when the replacement is manual, the volume is introduced rapidly or close to the dosage form, the agitation speed is low, or the formulation is particularly sensitive to agitation.
This is why non-replenishment with an appropriate correction factor is often the less intrusive approach.
The percentage released at each sampling time is calculated using the measured concentration and the medium volume actually present at that time. The calculation accounts for the progressive reduction in volume caused by earlier sampling.
But this approach has an important boundary. Consider a test conducted in 500 mL: If 5 mL is withdrawn at each of eight sampling time points, the final withdrawal represents a cumulative removal of 40 mL - 8% of the initial medium volume.
The decreasing volume can influence the hydrodynamic energy and mixing pattern. It can also affect the available sink margin of a poorly soluble drug operating close to saturation.
This is particularly important when the method is already operating under borderline sink conditions.
So which approach is better?
Neither should be selected as a default laboratory approach.
Non-replenishment avoids disturbing the vessel hydrodynamics and is generally preferable when:
- the cumulative withdrawal is small relative to the vessel volume
- sink conditions remain comfortably maintained
- the medium level and sampling zone remain suitable
Replenishment may become necessary when:
- a low medium volume is being used
- frequent sampling materially reduces the medium volume
- the drug is operating close to its solubility limit and reduction in volume could compromise sink conditions or method performance
But where replenishment is used, the addition procedure itself should be critically controlled.
Either ways, the selected approach must be predefined, scientifically justified and consistently executed.